Error Code |
Error Description |
Description |
Action |
21 |
Reference target |
Reference target statistics error, variation too high |
Check alignment of target / Return for Repair |
22 |
Reference target |
Reference target averaged value distance measurement too high |
Check alignment of target / Return for Repair |
23 |
Reference target |
Reference target - echo amplitude too small |
Check alignment of target / Return for Repair |
24 |
Reference target |
Reference target - echo amplitude too high |
Check alignment of target / Return for Repair |
25 |
Reference target |
Reference target- values occasionaly out of range |
Check for contamination / check alignment of target / Return for Repair |
26 |
Reference target |
Internal laser power measurement occasional out of limits |
Check reference target / Return for Repair |
27 |
Reference target |
Internal laser power measurement too small |
Check contamination / check alignment of target / Return for Repair |
28 |
Reference target |
Internal laser power measurement too large |
Check ailgnment of target / Return for Repair |
2A |
Reference target |
Reference target: distance too high |
Check alignment / return for Repair |
2B |
Reference target |
Reference target- echo amplitude too high |
Check alignment / return for Repair |
2C |
Reference target |
Reference target- distance too small |
Check alignment / return for Repair |
2D |
Reference target |
Reference target- echo amplitude too low |
Check alignment / return for Repair |
2F |
Reference target |
In Eeprom memorized reference target data defect |
Return for repair |
31 |
Trigger |
Rotating frequency too high |
Return for repair |
32 |
Trigger |
Rotating frequency too low |
Check whether head is blocked / Return for Repair |
33 |
Trigger |
Internal trigger signal missing |
Return for repair |
34 |
Trigger |
Internal trigger signal faulty |
Return for repair |
35 |
Trigger |
Internal synchronisation channel 1/2 not OK |
Return for repair |
41 |
Comparison |
Fault in comparison channel 1/2 in distance or echo amplitude |
Return for repair |
51 |
Relay |
Fault in relay monitoring |
Return for repair |
61 |
System |
Fault on self testing of CPU command |
Return for repair |
62 |
System |
Fault on self test RAM |
Return for repair |
63 |
System |
Fault on self test ROM |
Return for repair |
64 |
System |
Data error in internal Eeprom data |
Return for repair |
65 |
System |
Data error in Eeprom about Alarm or info parameters (User) |
Return for repair |
66 |
System |
Data error in Eeprom about Safe or Setup parameters (User) |
Return for repair |
67 |
System |
Data error in Eeprom in zone memory Return for repair |
Return for repair |
68 |
System |
Fault runtime, unexpected interrupt |
Return for repair |
69 |
System |
Fault runtime: fault in self testing sequence |
Return for repair |
70 |
Setup |
Last zone programming was unexpectedly broken off |
Reprogam the unit / Return for repair |